Comments on: Understanding SpO2 Sensor Testing https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/ 24x7 Magazine offers in-depth coverage and the latest news in Healthcare Technology Management, serving as the premier resource for HTM professionals seeking industry insights and updates. Mon, 31 Jan 2022 11:51:47 +0000 hourly 1 https://wordpress.org/?v=6.8.1 By: Geoff Mathews https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-156111 Sun, 22 Mar 2015 16:46:32 +0000 http://www.24x7mag.com/?p=30321#comment-156111 What are the moral and legal consequences when a patient is compromised by an inaccurate sensor?

Pulse oximeter sensor accuracy does matter. We see evidence of the impact of inaccurate plus oximeter sensors in mortality reviews where health care professionals have not known which sensor to believe.

What would happen if health care professionals and patients really knew what was going on with respect to inaccurate sensors?

Would simulator manufacturers that have failed to clearly advise that their products do not test sensor accuracy be liable?

]]>
By: Dr Douglas Clarkson https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-152420 Thu, 12 Mar 2015 17:31:51 +0000 http://www.24x7mag.com/?p=30321#comment-152420 The technology of pulse oximeters appears relatively simple but is in fact somewhat complex. The spectral output of the red and the infra red LEDs in the sensor unit is central to the clinical accuracy of the device and where an initial verification of readings against blood measurements of oxygen saturation is undertaken for a highly specific spectral output of each LED. If during manufacture or device use the output spectra (particularly of the red LED) diverge from the ideal specification, then the indicated patient SpO2 value will not be accurate. This can give rise to significant errors where, for example, an indicated value of 95% may in fact be 88%. This type of error can only be detected with a specialist device – such as the Lightman – which measures the spectral output of each LED and can compare this with the core specification of the specific pulse oximeter device.

]]>
By: Paul Horn https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-148670 Sat, 28 Feb 2015 19:55:06 +0000 http://www.24x7mag.com/?p=30321#comment-148670 I read this article with a mixture of interest, disappointment, and concern.

It provides a succinct, easily read and understood description of the basic principles. Unfortunately it misleads the reader by asserting that accuracy is solely a function of design, faithfully reproduced at manufacture, and that subsequent verification is neither possible, nor needed.

In practice, and in respect of sensor manufacture, this in fact can be far from the reality. In addition, there seems to be evidence (from spectral testing) that suggests an ‘ageing’effect, particularly with IR emitters over time. Any deviation from the actual wavelengths used during controlled desaturation studies to derive the original R curve, whether at manufacture, or in subsequent use, will alter that R curve, and will bias the accuracy of the pulse oximeter. A difference of just a few nanometres in the Red emitter wavelength is sufficient to cause inaccuracy beyond the standards requirement of 3% (and without this error being further compounded by biological variability).

The article acknowledges that functional testers have a role; they can be an aid to verify that any given system is functioning; they cannot verify that it is providing accurate readings, because they have no means of analysing/verifying the emitted wavelengths. Functional testing is important to ensure that monitors are working correctly (alarms etc), but this does absolutely nothing for establishing confidence in accuracy.

It is disappointing that this article has not made this distinction, and further confuses the reader by requiring a verification of accuracy of 3% in the performance test sheet (The Medical Equipment QA book linked to this article, Page 124). Whilst the requirement is correct, the functional testers described cannot verify this.

Verified functionality does not imply performance accuracy.

The publications and studies alluded to in the replies to this article have collectively sampled a large number of sensors, both new and used. The many errors identified would suggest that we should be at least as concerned with measuring sensor wavelength accuracy, as with verifying instrument function, and yet there is no acknowledgment of this as a test requirement.

Pulse oximetry at its inception in the 1980’s was hailed as “the most significant technological advance ever made in monitoring the well being of patients during anaesthesia, recovery and critical care”.
Today, we are more dependent (in the UK at least) than ever upon accuracy, since our Oxygen therapy policies are now predicated upon treating to a pulse oximeter saturation reading for diagnosis and treatment.

If the essential requirements of the relevant standards are to be met (accuracy to within 3%), and patient safety to be maintained, then accurate RED/IR wavelengths are central to achieving this, and the sooner agreement is reached upon the methods of ensuring it, the better.

]]>
By: Dr Geoff Mathews https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-146831 Tue, 17 Feb 2015 20:13:48 +0000 http://www.24x7mag.com/?p=30321#comment-146831 In reply to Fluke Biomedical.

Standards are often written with the benefit of hindsight available at the time they were written. Standards are the lowest common denominator that we should work to, but not be needlessly restricted by.

There appears to be some confusion over the process where the calibration data known as the R curve is established during the design phase, and the checking of accuracy of replacement sensors. The R curve is established by invasive desaturation studies where the results from a sensor of particular spectral properties are related to co-oximeter data. For replacement sensor to be accurate the emitter wavelengths must be a close match to the sensor used to establish the original R curve. As stated in the Pulse oximetry Standard (ISO 80601-2-61) – sensor accuracy is the match between the centre wavelengths emitted by the sensor and the calibration code used with that sensor.

The Fluke document referenced here is a long version of the list in the original article, and does not consider the impact and effect of replacement sensors. Sensor accuracy is important in pulse oximeter’s role in preventing adverse patient events. Users should be made aware that Healthcare Technology Management personnel following this referenced protocol do not test the correct match (accuracy) of the sensors, and that after an Annual Check of a monitor that the risk of a significant inaccuracy due to the sensor remains. Published data suggests that 30% of the sensors in use are unacceptable when tested. The cost benefit of appropriate clinical management of a patient and the cost risk of mismanagement needs to be remembered – this is a cost risk to a healthcare provider and not just the cost to an Engineering Department.

]]>
By: Fluke Biomedical https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-146820 Tue, 17 Feb 2015 18:01:02 +0000 http://www.24x7mag.com/?p=30321#comment-146820 Thank you all for feedback. This article touches on the limitations of testing pulse oximetry as a whole. We agree that the use of portable spectrometry could enhance fault detection of SPO2 sensors, and we welcome new innovation in this space. As with the medical technology itself, there are competing interests in ease of use, efficacy, and cost that are considered as new test technology is developed and deployed.

It should also be noted the IEC standard 80601-2-61 states, “There is today no accepted method of verifying the correct calibration of Pulse Oximeter Probe and Pulse Oximeter Monitor combination other than testing on human beings” (2014: line 2500).

Due to space limitations, this article does not discuss testing methods. We’d like to invite you download the full “Medical Equipment Quality Assurance” book referenced in this article to learn more about pulse oximetry test procedures and methods.

http://support.fluke.com/biomedical/Download/Asset/3276553_7150_ENG_A_W.PDF (Page 123)

]]>
By: Rhianwen Mathews (final year medical student) https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-146816 Tue, 17 Feb 2015 17:15:06 +0000 http://www.24x7mag.com/?p=30321#comment-146816 I recently conducted an audit looking at sensor accuracy of 227 pulse oximeter sensors over three different UK hospital trusts. I borrowed the Lightman from the UK based electrode company to run the accuracy tests. The Lightman is different from ‘functional’ testers as it measures the wavelength of the LED and compares this to the expected wavelength of the LED. If the LED puts out the wrong wavelength of light, a different amount of this light is absorbed by the blood, so a different amount of light passes through the finger to be read by the sensor, resulting in an inaccurate SpO2 measurement being calculated.

The results I found were really rather sobering.
In hospital one ~3% of the sensors were classed as unacceptable (unacceptable = error > +/- 3%).
In hospital two and three over one third of the sensors tested were classed as unacceptable.

Hospital one was unusual in that all of its sensors were of one make, which seems to have a much lower proportion of inaccurate sensors.

Scarily, the sensor with the highest bias of +8% was found on a patient in theatre. This patient’s true O2 saturation could have been significantly lower than that displayed on the monitor, with the potential for hypoxic damage.
The sensor with the largest low bias (-7%) was found in an anaesthetics room within the same hospital trust.

Depending which of these sensors was applied to the patient there could be a wide variety in what the sensors may read. For example:
– A patient with a true saturation of 90% could have a SpO2 value displayed of 94% (+/-3%) or 86% (+/-3%).
– With a SpO2 display of 90% the patient could have a true saturation of 84%(+/-3%) or 92%(+/-3%)

It is rather worrying that our hospitals have such a high proportion of inaccurate sensors! We need to start testing them for accuracy in addition to functionality!

]]>
By: Geoff Mathews https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-146797 Tue, 17 Feb 2015 14:26:31 +0000 http://www.24x7mag.com/?p=30321#comment-146797 In reply to mohammed niyaz ullah.

Sensor accuracy is dependent on placing LEDs of the correct wavelength in the sensor during production, and minimal changes in the spectral properties of the LEDs post production. This does not always happen and sensors can be wrong from new and can age.

The monitor is totally reliant on the red and IR wavelengths emitted by the sensor being correct. If these values are not correct the SATs values displayed on the monitor will not be correct. The SATs errors are greater at lower oxygen saturation values.

A short demonstration of a high reading sensor being tested can be viewed at: https://www.youtube.com/watch?v=pNxIAD_a_50

This error is mainly due to an error in the wavelength from the red LED. In use this sensor could lead to oxygen therapy and other important interventions being deferred or not given at all. No simulator or functional tester is capable of detecting this fault.

]]>
By: Dr Hickson https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-146796 Tue, 17 Feb 2015 14:10:07 +0000 http://www.24x7mag.com/?p=30321#comment-146796 Patrick O’Meley’s findings in Australia agree with my experience. More sensors have a low reading bias than high. This is not only of obvious importance in managing premature babies but also for COPD/COAD patients in whom a higher level of oxygen may lead to metabolic acidosis due to Carbon Dioxide retention and subsequent morbidity (and possible mortality).
A high biased sensor will lead to clinical management that could cause hypoxia to be prolonged with associated tissue damage and outcomes – for the neonate this could be cerebral palsy. Adults are not immune to lack of oxygen with outcomes ranging from confusion, strokes and death.
None of the workflow proposed by Andrew Clay i.e. “checking the physical condition, performing electrical safety tests, performing any manufacturer-recommended preventative maintenance, conducting performance testing (including alarms and other specific tests), and, finally, documenting the testing results” includes any check of accuracy. The pulse oximeter will only function as the manufacturer intended if the attached sensor has the correct colour LEDs for the R-curve to which it is calibrated. Testing the sensor for accuracy must be an essential part of any testing regime, otherwise it is falsely (and dangerously) reassuring.

]]>
By: mohammed niyaz ullah https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-146344 Sat, 14 Feb 2015 16:09:04 +0000 http://www.24x7mag.com/?p=30321#comment-146344 Its still not clear if the color pigments chanage due some reasons then what and how one has to rectifies it results

]]>
By: John Amoore https://24x7mag.com/medical-equipment/testing-calibration/general-testing-equipment/understanding-spo2-sensor-testing/#comment-146313 Sat, 14 Feb 2015 12:03:26 +0000 http://www.24x7mag.com/?p=30321#comment-146313 Andrew Clay’s article is welcome in that it highlights that pulse oximetry has become an accepted and recognised standard of patient care.
Where the article could be misleading is how it addresses the testing of pulse oximeters. The article notes that verification (type testing at design/factory stage) is carried out through controlled desaturation studies – not suitable for testing by biomeds.
The article goes on to quote Tobey Clarke’s recommendation of annual functional testings – leaving unclear how this can/should be carried out.

My concern with the article is that it does not really emphasise:
a. Limitations of simulators: that all that virtually all the spo2 simulators on the market can perform is indicating that the device is working, not that it is working accurately. Indeed, it could be argued that the simulator does little more than can be achieved by the biomed using the device on his/her own finger

b. The importance of accuracy, as other have commented on

c. That accuracy depends on the integrity of the sensor and in particular the wavelenghts of the light emmitted by the red and infrared LEDS. And that changes in the wavelenght from the design specification (poor batch, change in wavelenght with operating conditions) are likely to affect the accuracy

d. That methods are available to check the wavelenght of the LED lights sources

Best wishes

John

]]>